Ngwa: Mgbakọ tube X-ray maka nyocha nyocha nke oge niile na nke nkịtị
ma ọ bụ redio dijitalụ na ụlọ ọrụ fluoroscopic
Atụmatụ ntinye: 12.5º Rhenium-Tungsten molybdenum lekwasịrị anya (RTM)
◆Ntụgharị uche: Obere 0.6, Nnukwu: 1.2
◆ Oke voltaji tube: 125kV
◆ Ejiri ụdị IEC60526 nweta ihe nkwụnye ọkụ dị elu.
◆High voltaji generator kwesịrị kwekọọ na IEC60601-2-7
◆ IEC Nhazi (IEC 60601-1: 2005): Klas I ME Equipment

Nkọwa ngwaahịa

Usoro ịkwụ ụgwọ & Mbupu:

Mkpado ngwaahịa

Ntuziaka ọkọlọtọ

Ntuziaka ọkọlọtọ

Aha

EN 60601-2-54:2009 Ngwa eletriki ahụike - Nkebi 2-54: Ihe ndị achọrọ maka nchekwa bụ isi na arụmọrụ dị mkpa nke akụrụngwa X-ray maka redio na redio
IEC60526 Igwe eriri igwe dị elu na njikọ oghere maka akụrụngwa X-ray ahụike
IEC 60522:1999 Mkpebi nke nzacha na-adịgide adịgide nke mgbakọ tube X-ray
IEC 60613-2010 Njirimara eletriki, thermal na loading nke tubes X-ray anode na-atụgharị maka nyocha ahụike
IEC60601-1: 2006 Akụrụngwa eletriki ahụike - Nkebi 1: Ihe achọrọ maka nchekwa bụ isi na arụmọrụ dị mkpa
IEC 60601-1-3: 2008 Ngwa eletriki ahụike - Nkebi 1-3: Ihe ndị achọrọ maka nchekwa bụ isi na arụmọrụ dị mkpa
IEC60601-2-28:2010 Ngwa eletriki ahụike - Nkebi 2-28: Ihe ndị achọrọ maka nchekwa bụ isi na arụmọrụ dị mkpa nke mgbakọ X-ray tube maka nyocha ahụike.
IEC 60336-2005 Akụrụngwa eletriki ahụike- Mgbakọ tube X-ray maka nyocha ahụike-Ụdị ebe ntụpọ

Nkọwa

●Edebere aha ya dị ka ndị a:

MWHX7110A

Tube

A

Igwe ọkụ dị elu nwere ntụzịaka ogo 90

MWTX71-0.6/1.2-125

B

Igwe ọkụ dị elu nwere ntụzịaka ogo 270

Data teknụzụ

Ngwongwo

Nkọwapụta

Ọkọlọtọ

Ike ntinye aha nke anode

F 1

F 2

IEC 60613

20kW (50/60Hz)

40kW (50/60Hz)

 

Ike nchekwa okpomọkụ anode

110kJ (150kHU)

IEC 60613

Ike kachasị mma nke anode

500W

 
Ike nchekwa ọkụ

900kJ

 
Oke. okpomọkụ na-aga n'ihu na-enweghị ikuku- okirikiri

180W

 
Ihe anodeIhe mkpuchi elu anode

Rhenium-Tungsten-TZM(RTM)

Rhenium-Tungsten (RT)

 
Akụkụ ebumnuche (Ntụgharị: axis ntụaka)

12,5 Celsius

IEC 60788

Mgbakọ tube X-ray filtration pụtara

1.5 mm Al / 75kV

IEC 60601-1-3

Uru aha (s) n'ebe ntụpọ

F1 (obere elekwasị anya)

F2 (nnukwu uche)

IEC 60336

0.6

1.2

 
X-ray tube nominal voltajiRadiographicFluoroscopic

125kV

100kV

IEC 60613

Data na kpo oku cathode Oke. ugbu a

Oke voltaji

≈ / AC, <20 kHz

 

F1

F 2

 

5.1A

≈79V

5.1 A

≈1214 V

 
Nhapu radieshon na 150 kV/3mA na anya 1m

0.5mGy/h

IEC60601-1-3

Mpaghara radieshon kacha

443×443mm na SID 1m

 
X-ray tube mgbakọ arọ

Ihe ruru. 18 n'arọ

 

Ọnọdụ maka ọrụ, nchekwa na njem

Oke

Oke ọrụ

Oke njem na nchekwa

Ọnọdụ okpomọkụ

Site na 10ruo 40

Site na -20to 70

Iru mmiri dị ala

≤75%

≤93%

Mmetụta barometric

Site na 70kPa ruo 106kPa

Site na 70kPa ruo 106kPa

 

Ụkpụrụ igodo Stator

1-phase stator

Ebe ule

C-M

C-A

Nguzogide ikuku

≈18.0…22.0Ω

≈45.0…55.0Ω

Voltaji na-arụ ọrụ nke ukwuu (agba ọsọ)

230V± 10%

Kwado voltaji na-arụ ọrụ(gbalite)

160V± 10%

voltaji breeki

70VDC

Na-agba ọsọ voltaji na ikpughe

80Vrms

Voltaji na-agba ọsọ na fluoroscopy

20V-40Vrms

Oge ịgba ọsọ (dabere na usoro mmalite)

1.2s

Ịkpachara anya na arụ ọrụ

1 .X-ray Radiationnchebe

Ngwaahịa a na-emezu ihe achọrọ nke IEC 60601-1-3.

Mgbakọ tube X-ray a na-ewepụta radieshon X-ray na-arụ ọrụ. Naanị ndị ọrụ ruru eru na ndị zụrụ azụ ka a na-ahapụ ka ha rụọ mgbakọ tube X-ray.

Mmetụta physiologic dị mkpa nwere ike imerụ onye ọrịa ahụ, nrụpụta sistemụ kwesịrị ịnwe nchebe kwesịrị ekwesị iji zere radieshon ionization.

2.Dielectric 0il

Mgbakọ tube X-ray nwere dielectric 0il dị maka nkwụsi ike dị elu. Dị ka ọ na-egbu egbu maka ahụ ike mmadụ,ọ bụrụ na ekpughere ya na mpaghara enweghị mmachi,a ga-atụfu ya dịka usoro iwu obodo.

3 .Operation Atmosphere

A naghị ekwe ka e jiri mgbakọ x-ray tube mee ihe na ikuku gas na-ere ọkụ ma ọ bụ nke na-emebi emebi ·

4.Gbanwee tube ugbu a

Dabere na ọnọdụ ọrụ,enwere ike gbanwee njirimara filament.

Mgbanwe a nwere ike imetụta mkpughe nke oke ọnụ na mgbakọ tube X-ray.

Iji gbochie mgbakọ tube X-ray mebiri emebi,gbanwee tube ugbu a mgbe niile.

E wezụga mgbe X-ray tube nwere arcing Nsogbu na aleji oge eme ihe,chọrọ mgbanwe nke tube ugbu a.

5.X-ray Tube Okpomọkụ Ebe obibi

Emetụkwala n'elu ụlọ ụlọ X-ray tube obere oge ka arụchara ọrụ n'ihi oke okpomọkụ.

Nọgidere tube X-ray ka ọ dị jụụ.

6Oke ọrụ

Tupu eji,biko gosi na ọnọdụ gburugburu ebe obibi dị n'ime Imitts na-arụ ọrụ.

7.Arụ ọrụ ọ bụla

P1ease kpọtụrụ SAILRAY ozugbo,ọ bụrụ na achọpụtara adịghị arụ ọrụ ọ bụla nke mgbakọ tube X-ray.

8.Mkpofu

The X-ray tube mgbakọ nakwa dị ka tube nwere ihe ndị dị ka mmanụ na arọ ọla nke na gburugburu ebe obibi enyi na enyi na kwesịrị ekwesị mkpofu na dị na nti mba iwu ụkpụrụ ga-emesi.Ntufu dị ka anụ ụlọ ma ọ bụ ulo oru iju bụ forbidden.The emeputa nwere chọrọ oru ihe ọmụma na ga-ewe X-ray tube mgbakọ azụ maka mkpofu.

Biko kpọtụrụ ndị ahịa maka ebumnuche a.

Emission akụkụ nke cathode

图片

Ọ bụrụ (A) Obere ntụpọ

图片1.2焦点

Ọ bụrụ (A) Nnukwu Focal Spot

Ibu otu na usoro

Ọnọdụ: Tube Voltage Nkeji atọ

Ugboro ike Stator 50Hz/60Hz

Ia (mA)

图片4

t(s)

Ia (mA)

图片3

t(s)

Kpo oku na jụrụ usoro nke anode

  IEC60613

图片5

Kpọọ ọkụ na nju oyi nke mgbakọ tube X-ray

Njirimara okpomọkụ nke ụlọ

图片6

X-ray tube mgbakọ akụkụ eserese

SRMWHX7110A

图片7

Mgbakọ nzacha na ngalaba Cross Port

图片 27

Rotor Njikọ Wiring

图片 28

  • Nke gara aga:
  • Osote:

  • Ọnụ ego kacha nta: 1pc

    Ahịa: mkparita uka

    Nkọwa ngwugwu: 100pcs kwa katọn ma ọ bụ ahaziri dị ka ọnụọgụgụ

    Oge nnyefe: 1 ~ 2 izu dị ka ọnụọgụgụ

    Usoro ịkwụ ụgwọ: 100% T / T n'ihu ma ọ bụ WESTERN UNION

    Ịkwanye ikike: 1000pcs / ọnwa