Emepụtara ma mepụta ngwaahịa a na nkwekọrịta na iwu, ntụzịaka na ụkpụrụ imewe:
◆ Ntuziaka Council 93/42/EEC nke 14 June 1993 gbasara ngwaọrụ ahụike(Akara CE).
TS EN ISO 13485 Ngwa ahụike - Sistemụ njikwa ogo - Ihe achọrọ maka usoro iwu.
ebumnuche..
TS EN ISO 14971 Ngwa ahụike - Ngwa njikwa ihe egwu na ngwaọrụ ahụike
TS EN ISO 15223-1 ngwaọrụ ahụike - akara iji jiri akara ngwaọrụ ahụike, akara na ozi a ga-enye - Nkebi 1: Ihe ndị chọrọ n'ozuzu
◆International Electrotechnical Commission (IEC), a na-atụle ụkpụrụ ndị a karịsịa.
Ntuziaka ọkọlọtọ | Aha |
EN 60601-2-54:2009 | Ngwa eletriki ahụike - Nkebi 2-54: Ihe ndị achọrọ maka nchekwa bụ isi na arụmọrụ dị mkpa nke akụrụngwa X-ray maka redio na redio |
IEC60526 | Igwe eriri igwe dị elu na njikọ oghere maka akụrụngwa X-ray ahụike |
IEC 60522:1999 | Mkpebi nke nzacha na-adịgide adịgide nke mgbakọ tube X-ray |
IEC 60613-2010 | Njirimara eletriki, thermal na loading nke tubes X-ray anode na-atụgharị maka nyocha ahụike |
IEC60601-1: 2006 | Akụrụngwa eletriki ahụike - Nkebi 1: Ihe achọrọ maka nchekwa bụ isi na arụmọrụ dị mkpa |
IEC 60601-1-3: 2008 | Ngwa eletriki ahụike - Nkebi 1-3: Ihe ndị achọrọ maka nchekwa bụ isi na arụmọrụ dị mkpa |
IEC60601-2-28:2010 | Ngwa eletriki ahụike - Nkebi 2-28: Ihe ndị achọrọ maka nchekwa bụ isi na arụmọrụ dị mkpa nke mgbakọ X-ray tube maka nyocha ahụike. |
IEC 60336-2005 | Akụrụngwa eletriki ahụike- Mgbakọ tube X-ray maka nyocha ahụike-Ụdị ebe ntụpọ |
●Edebere aha ya dị ka ndị a:
MWHX7010 | Tube | A | Igwe ọkụ dị elu nwere ntụzịaka ogo 90 |
MWTX70-1.0/2.0-125 | B | Igwe ọkụ dị elu nwere ntụzịaka ogo 270 |
Ngwongwo | Nkọwapụta | Ọkọlọtọ | |
Ike ntinye aha nke anode | F 1 | F 2 | IEC 60613 |
21kW (50/60Hz) | 42.5kW (50/60Hz) | ||
Ike nchekwa okpomọkụ anode | 100kJ (140kHU) | IEC 60613 | |
Ike kachasị mma nke anode | 475W | ||
Ike nchekwa ọkụ | 900kJ | ||
Oke. okpomọkụ na-aga n'ihu na-enweghị ikuku- okirikiri | 180W | ||
Ihe anodeIhe mkpuchi elu anode | Rhenium-Tungsten-TZM(RTM) Rhenium-Tungsten (RT) | ||
Akụkụ ebumnuche (Ntụgharị: axis ntụaka) | 16 Celsius | IEC 60788 | |
Mgbakọ tube X-ray filtration pụtara | 1.5 mm Al / 75kV | IEC 60601-1-3 | |
Uru aha (s) n'ebe ntụpọ | F1 (obere elekwasị anya) | F2 (nnukwu uche) | IEC 60336 |
1.0 | 2.0 | ||
X-ray tube nominal voltajiRadiographic Fluoroscopic | 125kV 100kV | IEC 60613 | |
Data na kpo oku cathode Oke. ugbu a Oke voltaji | ≈ / AC, <20 kHz | ||
F1 | F 2 | ||
5.1A ≈5.8~7.8V | 5.1 A ≈7.7~10.4 V | ||
Nhapu radieshon na 150 kV/3mA na anya 1m | ≤1.0mGy/h | IEC60601-1-3 | |
Mpaghara radieshon kacha | 573×573mm na SID 1m | ||
X-ray tube mgbakọ arọ | Ihe ruru. 18 n'arọ |
Oke | Oke ọrụ | Oke njem na nchekwa |
Ọnọdụ okpomọkụ | Site na 10℃ruo 40℃ | Site na -20℃to 70℃ |
Iru mmiri dị ala | ≤75% | ≤93% |
Mmetụta barometric | Site na 70kPa ruo 106kPa | Site na 70kPa ruo 106kPa |
1-phase stator
Ebe ule | C-M | C-A |
Nguzogide ikuku | ≈18.0…22.0Ω | ≈45.0…55.0Ω |
Voltaji na-arụ ọrụ nke ukwuu (agba ọsọ) | 230V± 10% | |
Kwado voltaji na-arụ ọrụ(gbalite) | 160V± 10% | |
voltaji breeki | 70VDC | |
Na-agba ọsọ voltaji na ikpughe | 80Vrms | |
Voltaji na-agba ọsọ na fluoroscopy | 20V-40Vrms | |
Oge ịgba ọsọ (dabere na usoro mmalite) | 1.2s |
Ịdọ aka ná ntị iji jiri X-ray Generator mee ihe
1. Ụlọ mgbawa
Etinyekwala ike karịrị ike ọ bụla na mgbakọ tube X-ray
Ọ bụrụ na ike ntinye aka karịa nkọwa nke tube ahụ, ọ nwere ike ime ka anode na-ekpo ọkụ, iko tube na-agbaji, na n'ikpeazụ nwere ike ịkpata nsogbu siri ike n'ihi oke ọkụ na-akpata site na vaporization nke mmanụ n'ime mgbakọ ụlọ. N'ọnọdụ dị oke egwu ebe ụlọ na-agbaji n'ihi oke ibu, nchekwa ọkụ ọkụ nwere ike ọ gaghị echebe tube X-ray, ọ bụrụgodị na ọ na-arụ ọrụ.
*Ụlọ akara akụkụ mgbawa.
*Mmerụ mmadụ gụnyere ọkụ n'ihi mgbapụ mmanụ ọkụ.
*Ihe mberede ọkụ n'ihi ebumnuche anode na-ere ọkụ.
Igwe ọkụ X-ray kwesịrị inwe ọrụ nchekwa nke na-ejikwa ike ntinye ka ọ dịrị n'ime nkọwapụta tube.
2.Electric Ujo
Iji zere ihe egwu nke ujo eletrik, akụrụngwa a ga-ejikọrịrị na ọkọnọ nwere ụwa nchebe.
3.Ọ dịghị mgbanwe nke ngwá ọrụ a na-ekwe !!
Kpachara anya ka ị jiri X-ray Generator mee ihe
1. Over Rating
Ike dị ukwuu n'otu égbè nwere ike ime ka mgbakọ X-ray tube ghara ịdaba. Ọ dị mkpa iji nlezianya nyochaa akwụkwọ data teknụzụ ma soro ntuziaka akọwapụtara iji zere mmebi.
2.Nzacha na-adịgide adịgide
Usoro iwu na-akọwapụta mkpokọta nzacha achọrọ yana anya kacha nta n'etiti ebe a na-elekwasị anya X-ray na ahụ mmadụ.
They kwesịrị irubere iwu.
3.Nchekwa ọkụ ọkụ
Mgbakọ tube X-ray nwere mgba ọkụ ọkụ na-echebe iji gbochie ike ntinye ọzọ mgbe ụlọ tube ruru na okpomọkụ(80℃)nke mgba ọkụ-emeghe.
A naghị atụ aro ka mgba ọkụ ahụ jikọọ eriri stator na sekit usoro.
Ọbụlagodi na mgba ọkụ na-arụ ọrụ, agbanyụla sistemu ike. Ekwesịrị ịgbalite nkeji jụrụ oyi ma ọ bụrụ na ejiri ya na sistemụ.
4.Atụghị anya malfunction
Mgbakọ tube X-ray nwere ike ịrụ ọrụ nke ọma ma ọ bụ daa na mberede, na-ebute nsogbu nke nnukwu nsogbu. Ọ dị mkpa ịnwe atụmatụ nchekwa iji gbochie na dozie nsogbu ọ bụla nwere ike ibilite na ihe egwu a.
5. Ngwa ọhụrụ
Ọ bụrụ na ị na-eme atụmatụ iji ngwaahịa a na ngwa ọhụrụ akọwapụtaghị n'ime akwụkwọ a, ma ọ bụ ọ bụrụ na ị na-eme atụmatụ iji ụdị generator X-ray dị iche, biko kpọtụrụ anyị iji kwado ndakọrịta na ịdị adị.
1 .X-ray Radiationnchebe
Ngwaahịa a na-emezu ihe achọrọ nke IEC 60601-1-3.
Mgbakọ tube X-ray a na-ewepụta radieshon X-ray na-arụ ọrụ. Naanị ndị ọrụ ruru eru na ndị zụrụ azụ ka a na-ahapụ ka ha rụọ mgbakọ tube X-ray.
Mmetụta physiologic dị mkpa nwere ike imerụ onye ọrịa ahụ, nrụpụta sistemụ kwesịrị ịnwe nchebe kwesịrị ekwesị iji zere radieshon ionization.
2.Dielectric 0il
Mgbakọ tube X-ray nwere dielectric 0il dị maka nkwụsi ike dị elu. Dị ka ọ na-egbu egbu maka ahụ ike mmadụ,ọ bụrụ na ekpughere ya na mpaghara enweghị mmachi,a ga-atụfu ya dịka usoro iwu obodo.
3 .Operation Atmosphere
A naghị ekwe ka e jiri mgbakọ x-ray tube mee ihe na ikuku gas na-ere ọkụ ma ọ bụ nke na-emebi emebi ·
4.Gbanwee tube ugbu a
Dabere na ọnọdụ ọrụ,enwere ike gbanwee njirimara filament.
Mgbanwe a nwere ike imetụta mkpughe nke oke ọnụ na mgbakọ tube X-ray.
Iji gbochie mgbakọ tube X-ray mebiri emebi,gbanwee tube ugbu a mgbe niile.
E wezụga mgbe X-ray tube nwere arcing Nsogbu na aleji oge eme ihe,chọrọ mgbanwe nke tube ugbu a.
5.X-ray Tube Okpomọkụ Ebe obibi
Emetụkwala n'elu ụlọ ụlọ X-ray tube obere oge ka arụchara ọrụ n'ihi oke okpomọkụ.
Nọgidere tube X-ray ka ọ dị jụụ.
6Oke ọrụ
Tupu eji,biko gosi na ọnọdụ gburugburu ebe obibi dị n'ime Imitts na-arụ ọrụ.
7.Arụ ọrụ ọ bụla
P1ease kpọtụrụ SAILRAY ozugbo,ọ bụrụ na achọpụtara adịghị arụ ọrụ ọ bụla nke mgbakọ tube X-ray.
8.Mkpofu
Mgbakọ tube X-ray yana tube nwere ihe ndị dị ka mmanụ na ọla dị arọ nke a ga-emesi obi ike na gburugburu ebe obibi enyi na ikpofu kwesịrị ekwesị dị ka ụkpụrụ iwu obodo siri dị. ihe ọmụma teknụzụ achọrọ ma ga-eweghachi mgbakọ tube X-ray maka mkpofu.
Biko kpọtụrụ ndị ahịa maka ebumnuche a.
Ọ bụrụ (A) Obere ntụpọ
Ọ bụrụ na (A) Nnukwu Ebe Focal
Ọnọdụ: Tube Voltage Nkeji atọ
Ugboro ike Stator 50Hz/60Hz
IEC60613
Njirimara okpomọkụ nke ụlọ
SRMWHX7010A
SRMWHX7010B
Mgbakọ nzacha na ngalaba Cross Port
Rotor Njikọ Wiring
Ọnụ ego kacha nta: 1pc
Ahịa: mkparita uka
Nkọwa ngwugwu: 100pcs kwa katọn ma ọ bụ ahaziri dị ka ọnụọgụgụ
Oge nnyefe: 1 ~ 2 izu dị ka ọnụọgụgụ
Usoro ịkwụ ụgwọ: 100% T / T n'ihu ma ọ bụ WESTERN UNION
Ịkwanye ikike: 1000pcs / ọnwa